Thailand: Change Notification System for Class 2–4 Medical Devices

The Thai Food and Drug Administration (Thai FDA) has published an infographic outlining the Change Notification System for Class 2–4 medical devices.

The update explains how changes are categorized, the applicable processing timelines, correction periods, and government and expert assessment fees. Processing requirements differ depending on whether the proposed change has already been approved by a reference country, the number and type of changes submitted, and whether external expert review is required.

Read the full update here:
https://medical.fda.moph.go.th/press-release/change-notification-system/

(Thai FDA official website. Information is provided in Thai.)

Levels of Change

Changes are divided into three categories:

Major and Minor Changes

Major and Minor changes require technical document assessment by a Thai FDA officer or external expert before approval.

Auto Approve

Certain eligible changes may be automatically approved through the system without further assessment.

Processing Time and Review Conditions

Approved Change Received from a Reference Country

CaseChange Topic in ApplicationProcessing Time Without ExpertProcessing Time With ExpertCorrection RoundsTime Given for Correction
1Adding items to the medical device list, up to 10 items; or other Major and Minor changes15 working days–15 working days
2Adding items to the medical device list, more than 10 items35 working days–15 working days

For this condition, “adding items to the medical device list” includes:

  • Adding a medical device item – Major
  • Adding a medical device item with the same design – Major
  • Adding a medical device item or software that does not affect performance characteristics or product specifications – Minor
  • Adding accessory items to the main medical device – Minor

Approved Change NOT Received from a Reference Country

CaseChange Topic in ApplicationReview TypeProcessing TimeCorrection Rounds by OfficerCorrection Rounds by ExpertTime Given by OfficerTime Given by Expert
1Auto Approve: 1 topic or more, provided the application contains only Auto Approve topicsAutomaticImmediate––––
2Major: 1 topic, or Minor: 1–3 topicsWithout expert15 working days1–5 days–
2Major: 1 topic, or Minor: 1–3 topicsWith expert30 working days115 days5 days
3Major: 2–3 topics, or Minor: 4–9 topicsWithout expert35 working days1–10 days–
3Major: 2–3 topics, or Minor: 4–9 topicsWith expert55 working days1110 days10 days
4Major: 4–5 topics, or Minor: 10–15 topicsWithout expert50 working days1–10 days–
4Major: 4–5 topics, or Minor: 10–15 topicsWith expert70 working days1110 days10 days
5Major: 5+ topics, or Minor: 16+ topicsWithout expert65 working days1–10 days–
5Major: 5+ topics, or Minor: 16+ topicsWith expert85 working days1110 days10 days

Examples provided by Thai FDA include:

  • Case 3: Minor 5 topics; or Minor 3 topics + Major 1 topic
  • Case 4: Minor 11 topics; or Minor 3 topics + Major 3 topics
  • Case 5: Minor 17 topics; or Minor 3 topics + Major 4 topics

For Cases 2–5, Thai FDA applies the equivalence of:

3 Minor changes = 1 Major change

Fees

CategoryApplication FeeExpert Document Assessment FeeTotal With Expert
Notified Certificate of Manufacturing – Class 2–3THB 500THB 19,200THB 19,700
Notified Certificate of Importation – Class 2–3THB 500THB 24,000THB 24,500
Manufacturing License – Class 4THB 500THB 24,000THB 24,500
Import License – Class 4THB 500THB 30,000THB 30,500

If expert review is not required, only the THB 500 application fee applies.

Implications to Clients

Any proposed change to a registered Class 2–4 medical device should be submitted for regulatory assessment before implementation.

The applicable processing timeline and fees will depend on:

  • Whether the change has already been approved by a reference country
  • Whether the change is classified as Major, Minor, or Auto Approve
  • The number of changes included in the application
  • Whether external expert review is required

Clients should also consider the impact of combining multiple changes within a single application, as a higher number of changes may significantly extend the review timeline, particularly where expert assessment is required.

Effective Date

Effective from August 14, 2026

For Inquiries

For inquiries or support regarding medical device regulatory requirements in Thailand, please contact sales@andamanmed.com.